Product Quick Links

Data Acquisition and Data Logger Product Screener
Find the right product for you
with our Product Screener

DATAQ Instruments Blog

Category Archives: General Data Logger

Help Us Out– New Engineering Unit Settings Dialog Box

Categories: General Data Acquisition, General Data Logger 18 Comments

Help Us Out

We’re looking for customers to test this new WinDaq software approach for defining engineering units and give us their opinions. Currently the new code is available only for starter kits, so anyone with a DI-194(RS), -154RS, -148U, -148U-SP, -158U, -158UP, -145, -149, or DI-155 qualifies. To sign up, please leave a comment on this blog post with a valid email address, and we’ll follow up with instructions. Note that the new code doesn’t yet replace the other engineering unit scaling options. So, if you hate the new approach you have time to tell us why and have us fix it before it does. Finally, we’ve tested this code and found it to work well, so we hope your evaluation applies only to look and feel. Of course, if you trip on a bug please let us know.

What follows should give you a good feel for why we’re making this change and why its better. Of course, feel free to post any questions you may have in the comments section.

Engineering Unit Scaling Background

The first thing that most data acquisition and data logger equipment users want to do after they acquire analog data is to scale it into meaningful units. To most, acquired data displayed in volts is next to meaningless. They need the ability to translate volts into meaningful measures, like PSI, Newtons, ft-lb, mmHg, etc. These are often referred to as engineering units (EU), and the process of converting voltages applied to the input into meaningful measures like these is referred to as engineering unit scaling. Of course, DATAQ Instruments’ popular WinDaq application supports this feature, but you’ve let us know that the procedure, as well as what we call it, needs improvement.

You talked. We listened.

We heard you. Based upon countless conversations with you,  it was clear that our current High/Low method as well as the Fixed method of converting to engineering units was neither intuitive nor easy to use. We also sensed that we confused our customer base through our use of the word calibration to refer to EU scaling. Many customers rightfully feared that they would change the actual calibration of the data acquisition unit if they used these features. None of this is good news for a feature that is so crucial to data acquisition and data logging operations.

A New Engineering Unit Scaling Dialog Box

We’ve responded with a fresh approach to engineering unit scaling that we hope you will view as a major step forward for WinDaq ease-of-use.

Engineering units settings dialog box

The New WinDaq Engineering Units Settings Dialog Box

Nearby you will see a graphic image of our new Engineering Unit Settings dialog box, which will eventually replace our current High/Low and Fixed dilaog box approaches. I hope at first glance that, aside from minor questions, it’s obvious how to use it. In the most basic applications, you know the relationship of volts to engineering units and you simply define that relationship using the supplied dialog boxes. Some examples:

  • 10V equals 500 PSI and 0V equals zero PSI.
    • Under the Volt column, you’d set Upper Level equal to “10″ and Lower Level equal to “0″
    • Under the EU column, you’d set Upper Level equal to “500″ and Lower Level equal to “0″
    • Set the EU Tag dialog box to “PSI”
    • Done!
  • Using a 250-Ohm resistor on a process current loop where the sensor’s output is 300 lbs at 20 mA and 0 lbs at 4 ma
    • Since the 250-Ohm resistor converts the 4-20 mA signal to 1-5 V…
    • Under the Volt column, you’d set Upper Level equal to “5″ and Lower Level equal to “1″
    • Under the EU column, you’d set Upper Level equal to “300″ and Lower Level equal to “0″
    • Set the EU Tag dialog box to “LBS”
    • Done!

Some Additional Features

As you study the new Engineering Units Settings dialog box you’ll notice some interesting items starting with the eyedropper icons adjacent to the Volt data entry boxes.

We’ve noticed that many WinDaq users simply do not know the relationship of volts output from the sensor versus engineering units, so they derive the relationship empirically. Using the example of a load cell with an unknown output versus pounds of applied force, they’ll apply no load, then measure and record the voltage output. Then they’ll measure and record the output with a known load applied. The two recorded voltage values versus applied load can be entered in the appropriate boxes to define the relationship.

The eyedroppers eliminate a couple steps. They allow you to sample the current value of the applied voltage when you click the icon, and have it automatically displayed in the Volt box for either the Upper or Lower value depending upon the applied load. Continuing with the example, you’d apply no load and click the Lower eyedropper so the software will acquire the lower voltage value and display it automatically in the Lower box for voltage. Then you’d enter zero for the lower EU value. Next, you’d apply your known load and do the same for the Upper values. Enter your EU Tag and you’re done. As you can see for this calibration method, using the eyedroppers saves time and greatly reduces the possibility of human error.

Finally, there’s the Set Zero button that allows you to instantly compensate for zero offsets from your measurement. How many times have you carefully calibrated a system only to see a slight offset in the result when you apply zero engineering units? It’s very common. You can apply zero engineering units to your sensor and then use the Set Zero feature to instantly force the offset to zero when you click the button.

Help Facility

In the final version that we release to everyone, an additional button labeled Help (not available in this beta version) will provide a dedicated help reference just for the Engineering Unit Settings dialog box.

 

Connecting to your Temperature Alert LAN-based Temperature and Humidity Data Logger

Categories: General Data Logger, Temperature and Humidity Data Loggers Leave a comment

The ‘Connection Settings’ tab allows you to choose a hard-wired Ethernet or WiFi connection.  You can configure the Temperature Alert data logger a for DHCP or static operation, and scan for wireless networks.

An Overview of the Temperature Alert LAN-based Temperature and Humidity Data Logger

Categories: General Data Logger, Temperature and Humidity Data Loggers Leave a comment

A built-in web server gives you access to all of the Temperature Alert’s impressive features. Use your favorite web browser to view the current temperature and humidity, set alarm levels and specify an email address where alerts and daily reports can be sent.

How To Eliminate Noise in Data Logger and Data Acquisition Measurements

Categories: General Data Acquisition, General Data Logger, Thermocouple Data Acquisition Leave a comment

(Note: This is a consolidated version of a six-part series on signal noise reduction in data logger and data acquisition systems.)

This happens almost daily: A customer contacts our support group to complain of noise on his measurement. Almost invariably a comparison is made between the output of his PC-based product and a connected DVM. The DVM shows a stable reading while the PC-connected measurement “is jumping all over the place.”

While there’s a slim chance that the PC-based instrument is to blame, that can only be true if the front end has been damaged in some manner. While not perfect, analog front ends on data acquisition and logger products are designed to inject only very small amounts of noise as a percentage of full scale measurement range.

What should you do if your setup exhibits a large amount of noise, so much so that it’s difficult to discern the true measured value? This series makes some suggestions that will help you identify and eliminate the problem. In this first part we’ll show you how to ensure that the instrument is functioning properly and is not the source of the noise. In subsequent posts we’ll dig progressively deeper into field wiring to expose some common problems that can manifest themselves.

So, is it the instrument, or not? Follow these steps to find out:

  1. Are you over reacting? Our software allows you to expand signal amplitude ranges to such a degree that a relatively small amount of noise can look overbearing. It’s a lot like cringing in horror as you look at a tick under a microscope. Read How To Put Signal Noise Into Perspective for guidance.
  2. Assuming that the noise you see is a significant percentage of the full scale range as determined by step (1), then your next move is to eliminate the instrument as the culprit. Disconnect all field wiring from it and simply short the analog inputs to common. I know. In the electronics world the word “short” implies something that’s very bad, but in this case it’s perfectly kosher. If your instrument has a single-ended configuration, short the (+) input to signal common. If it’s a differential configuration, connect the (+) input to the (-) and then connect those to signal common (if available, which it may not be for some differential configurations.) What you’re doing is establishing a firm zero volts on the front end of the channel so you can examine the instrument’s noise floor. In this configuration you should see no more than a few counts of noise around zero volts, dramatically and substantially less than you saw with a signal source connected, and roughly consistent with the manufacturer’s specification for the instrument’s noise figure. If you see anything else, then contact tech support for guidance because the chances are great that your instrument is defective.

Next time, we’ll delve into common field wiring issues that can and do produce more noise than a heavy metal concert.

 

 

Common Mode Voltages and Isolated Instruments

In the premier edition of this series I talked about how you should initially approach noise elimination. The first step is to ensure that noise is a significant percentage of the full scale range of the analog channel, and then I explained a simple test to ensure that the instrument is not the noise source. If you’ve gotten this far I can assume that signal noise is significant, and the instrument is not defective.

Your next step is to look up the isolation specification of the instrument. Note that if you are under the impression that an instrument with differential analog inputs means that it is also isolated, you need to read this application note before proceeding. Differential and isolated analog inputs are not synonymous, and unless the specifications for the instrument state clearly that its inputs are isolated it most likely is not, regardless of its single-ended and/or differential configuration options. This edition of the Noise Elimination series assumes that you have an isolated instrument. I’ll deal with non-isolated instruments in later editions, since the subject becomes more complex in this situation.

How To Measure Common Mode Voltage Magnitude

So, you have read its specifications and determined that you have an isolated instrument. That’s great because noise-inducing common mode voltages (CMVs) can be largely, but not completely ignored. To absolutely verify that a common mode voltage is not a problem start by answering these questions:

  1. Is your field measurement one that virtually guarantees that a CMV is present? Examples are exposed thermocouple junctions that are attached directly to a conductor (powered or not), or the measurement of AC or DC current using a current shunt.
  2. If you answered YES to (1), then determine the magnitude of the CMV. With all field wiring disconnected except power and/or computer interfaces, use a battery-powered multimeter to measure the voltage from the field-side shunt or exposed conductive surface to the instrument’s chassis ground (not its low-side signal input.) Do this on the meter’s AC and DC ranges. Jot the value down and compare it with the isolation breakdown specification for the instrument.

If the measured value exceeds this spec then you have found the source of your noise, and you’re fortunate that the instrument still functions. Your two options are to lower the CMV if possible, or replace the instrument with one that provides a higher isolation spec.

In the next edition I’ll discuss CMV frequency and how this can destroy the best instrument isolation barriers.

Effects of Common Mode Voltage Frequency

We’ve covered evaluating noise magnitude to see if noise elimination is worth the time and effort, and a procedure to test the instrument for proper operation. In part 2, the concept of a common mode voltage (CMV) was introduced, and how a CMV can disrupt the operation of even isolated instruments if it’s too large. Here I’ll continue the discussion of common mode voltages in isolated instruments, focusing not on its magnitude, but on its frequency.

Most of the CMV situations you’ll encounter revolve around 50/60 Hz power, or maybe 100/120 Hz in the presence of florescent lighting. These are low frequency CMVs that won’t interfere with the ability of an isolation barrier to perform. There are applications however that can wreak havoc on an isolation barrier; in the presence of high frequency noise generated by a welding machine or some other piece of industrial equipment, for example.

As the frequency of the CMV increases, the effectiveness of the isolation barrier decreases. This is due to the parasitic capacitance inherent to any non-optical isolation amplifier (parasitic capacitance). The formula used to calculate capacitive reactance (Xc) illustrates this relationship.

Think of capacitive reactance as the amplifier’s ability to resist noise. As the frequency (f) of the CMV rises, the isolation amplifier becomes less effective. At a certain threshold frequency the CMV is no longer blocked by the isolation barrier, but effectively coupled through it because of the barrier’s inherent parasitic capacitance.

 

 

Common Mode Rejection Ration (CMRR)

We’ve touched on common mode voltage and how its magnitude and frequency can degrade the performance of an isolation amplifier. Now we’ll discuss the differential amplifiers ability to reject common mode voltage, a.k.a. its Common-mode rejection ratio (CMRR).

In theory, with identical signals connected simultaneously to the (+) and (-) leads of a differential amplifier, the output should be 0V. In reality, some small voltage will always be present. This is an error voltage that will ride on top of the expected output. A differential amplifier’s CMRR is an indicator of just how much error voltage might be present on the output.

The common-mode rejection ratio is calculated as follows:

 

CMRR = 20 LOG(Vin(cm)/Vout) dB

Where:

Vin(cm) = the voltage common to both the + and – leads of the amplifier

 

So for example, assuming that you have 5V connected to the + and – leads of a differential amplifier and you see an output of 158µV, your CMRR is 90dB.

By arranging the formula, with 5V common to the + and – leads of the amplifier and a known CMRR of 100dB, we know that the output will be ±50µV:

Vout = VCM/10CMRR/20

As these examples illustrate, the higher the CMRR, the less common voltage is passed through to the output, resulting in a cleaner signal.

Proper Shielding Techniques

We’ve established that isolation amplifiers do not perform well in the presence of large, high frequency common-mode voltages. Now we’ll focus on eliminating high level, high frequency signals before they ever reach the amplifier.

When connecting signals to your data acquisition instrument, proper shielding can make all the difference in the world. This is especially true when dealing with low level signals, where a little noise can render your data useless.

The shielding inside of a shielded cable is made up of strands of braided aluminum or copper (and sometime foil) that surround the connecting wires, insulating them from environmental noise. A “drain” wire is used to terminate the shield to ground. It is connected on either the sensor OR the instrumentation end of the cable (never both as this could result in a ground loop). Shielding is effective at reducing noise in single-ended or differential applications.

Twisting signal wires together to form a twisted pair is another method used to reduce noise. This is done so that neither wire is exposed to more noise than the other. As a result any noise that does make it to the amplifier is identical on both the + and – leads, and is therefore canceled out.

Common Mode Voltages and Non-isolated Instruments

You’ve arrived at part six of this series because you have noise on one or more acquired analog channels. Using procedures described in Part 1 of the series you’ve determined that the noise is a significant percentage of the instrument’s full scale range and is not attributable to the instrument. Further, you have determined that your instrument is not isolated, or at least does not provide both input-to-output and channel-to-channel isolation. Given this last fact, it is almost certain that your noise is caused by a common mode voltage.

Whenever the ground reference of a non-isolated instrument is connected to the ground reference of a non-isolated signal source and the grounds are at different potentials, current will flow through the grounds and manifests itself as noise in the acquired signal. Some points to ponder:

  1. Common mode voltage can be defined as the potential difference between two grounds.
  2. The ground reference of the signal and the instrument are almost never at the same potential in industrial settings.
  3. The magnitude of noise will be directly proportional to the magnitude of the common mode voltage.
  4. The sum of common mode voltage plus signal amplitudes can exceed the maximum that your instrument will tolerate and cause permanent damage.
  5. If the signal source is isolated, then the instrument does not need to be and vice versa.

It should be clear from the above that the elimination of common mode voltages leads directly to the elimination of noise. How you do that will be different in almost every situation, but realizing that the term “ground” is strictly relative, and that any given set of grounds are almost always at different potentials is a great starting point. Follow these tips to eliminate noise-inducing common mode voltages:

  1. Examine the power supplies of your signal source and instrument. Are they plugged into the same outlet or terminal strip? If not then try that as a first remedy.
  2. Have you connected your instrument to the signal source using shielded cable? Is the shield connected to ground on both ends? If so, then remove the connection on one end to prevent current from flowing through your shield, destroying its effectiveness.
  3. If you’re using thermocouples, are the junctions in direct contact with a grounded, metallic object? If so, then consider using an insulating material between the junction and the metallic object that removes the CMV, and yet provides good thermal conductivity.
  4. Are you measuring across a shunt resistor, like a 250Ω resistor in a 4-20 mA process current application? If so, then ensure that the resistor is in stalled in the low side of the loop. Placing the resistor in the high side of the loop exposes your instrument to a common mode voltage equal to the loop power supply voltage.
  5. Even if you’ve placed the shunt resistor in the low side of the 4-20 mA current loop, make sure that the loop power supply’s ground is floating (isolated) from input power common. If it isn’t, common mode voltages that most likely exist between loop and instrument grounds will cause trouble.

If you’ve examined your field wiring for all of the above and still cannot solve the problem, then you may have to toss in the towel and move to an isolated instrument that is largely immune to the problems caused by a CMV. Depending upon your application, you may be surprised at how little isolation costs, how much time it saves you, and much your measurement improves.

 

 

How To Eliminate Noise in Data Logger and Data Acquisition Measurements (Part 6 of 6)

Categories: General Data Acquisition, General Data Logger, Thermocouple Data Acquisition Leave a comment

(Note: A consolidated version of this topic that contains all six part can be found here)

Common Mode Voltages and Non-isolated Instruments

You’ve arrived at part six of this series because you have noise on one or more acquired analog channels. Using procedures described in Part 1 of the series you’ve determined that the noise is a significant percentage of the instrument’s full scale range and is not attributable to the instrument. Further, you have determined that your instrument is not isolated, or at least does not provide both input-to-output and channel-to-channel isolation. Given this last fact, it is almost certain that your noise is caused by a common mode voltage.

Whenever the ground reference of a non-isolated instrument is connected to the ground reference of a non-isolated signal source and the grounds are at different potentials, current will flow through the grounds and manifests itself as noise in the acquired signal. Some points to ponder:

  1. Common mode voltage can be defined as the potential difference between two grounds.
  2. The ground reference of the signal and the instrument are almost never at the same potential in industrial settings.
  3. The magnitude of noise will be directly proportional to the magnitude of the common mode voltage.
  4. The sum of common mode voltage plus signal amplitudes can exceed the maximum that your instrument will tolerate and cause permanent damage.
  5. If the signal source is isolated, then the instrument does not need to be and vice versa.

It should be clear from the above that the elimination of common mode voltages leads directly to the elimination of noise. How you do that will be different in almost every situation, but realizing that the term “ground” is strictly relative, and that any given set of grounds are almost always at different potentials is a great starting point. Follow these tips to eliminate noise-inducing common mode voltages:

  1. Examine the power supplies of your signal source and instrument. Are they plugged into the same outlet or terminal strip? If not then try that as a first remedy.
  2. Have you connected your instrument to the signal source using shielded cable? Is the shield connected to ground on both ends? If so, then remove the connection  on one end to prevent current from flowing through your shield, destroying its effectiveness.
  3. If you’re using thermocouples, are the junctions in direct contact with a grounded, metallic object? If so, then consider using an insulating material between the junction and the metallic object that removes the CMV, and yet provides good thermal conductivity.
  4. Are you measuring across a shunt resistor, like a 250Ω resistor in a 4-20 mA process current application? If so, then ensure that the resistor is in stalled in the low side of the loop. Placing the resistor in the high side of the loop exposes your instrument to a common mode voltage equal to the loop power supply voltage.
  5. Even if you’ve placed the shunt resistor in the low side of the 4-20 mA current loop, make sure that the loop power supply’s ground is floating (isolated) from input power common. If it isn’t, common mode voltages that most likely exist between loop and instrument grounds will cause trouble.

If you’ve examined your field wiring for all of the above and still cannot solve the problem, then you may have to toss in the towel and move to an isolated instrument that is largely immune to the problems caused by a CMV. Depending upon your application, you may be surprised at how little isolation costs, how much time it saves you, and much your measurement improves.

 

 

 

 

How To Eliminate Noise in Data Logging and Data Acquisition Measurements (Part 5 of 6)

Categories: General Data Acquisition, General Data Logger, Thermocouple Data Acquisition Leave a comment

(Note: A consolidated version of this topic that contains all six part can be found here)

Proper Shielding Techniques

We’ve established that isolation amplifiers do not perform well in the presence of large, high frequency common-mode voltages. Now we’ll focus on eliminating high level, high frequency signals before they ever reach the amplifier.

When connecting signals to your data acquisition instrument, proper shielding can make all the difference in the world. This is especially true when dealing with low level signals, where a little noise can render your data useless.

The shielding inside of a shielded cable is made up of strands of braided aluminum or copper (and sometime foil) that surround the connecting wires, insulating them from environmental noise. A “drain” wire is used to terminate the shield to ground. It is connected on either the sensor OR the instrumentation end of the cable (never both as this could result in a ground loop). Shielding is effective at reducing noise in single-ended or differential applications.

Twisting signal wires together to form a twisted pair is another method used to reduce noise. This is done so that neither wire is exposed to more noise than the other. As a result any noise that does make it to the amplifier is identical on both the + and – leads, and is therefore canceled out.

 

How To Eliminate Noise in Data Logging and Data Acquisition Measurements (Part 4 of 6)

Categories: General Data Acquisition, General Data Logger, Thermocouple Data Acquisition Leave a comment

(Note: A consolidated version of this topic that contains all six part can be found here)

Common Mode Rejection Ration (CMRR)

We’ve touched on common mode voltage and how its magnitude and frequency can degrade the performance of an isolation amplifier. Now we’ll discuss the differential amplifiers ability to reject common mode voltage, a.k.a. its Common-mode rejection ratio (CMRR).

In theory, with identical signals connected simultaneously to the (+) and (-) leads of a differential amplifier, the output should be 0V. In reality, some small voltage will always be present. This is an error voltage that will ride on top of the expected output. A differential amplifier’s CMRR is an indicator of just how much error voltage might be present on the output.

The common-mode rejection ratio is calculated as follows:

 

CMRR = 20 LOG(Vin(cm)/Vout) dB

 Where:

Vin(cm) = the voltage common to both the + and – leads of the amplifier

 

So for example, assuming that you have 5V connected to the + and – leads of a differential amplifier and you see an output of 158µV, your CMRR is 90dB.

By arranging the formula, with 5V common to the + and – leads of the amplifier and a known CMRR of 100dB, we know that the output will be ±50µV:

Vout = VCM/10CMRR/20

 As these examples illustrate, the higher the CMRR, the less common voltage is passed through to the output, resulting in a cleaner signal.

How To Eliminate Noise in Data Logger and Data Acquisition Measurements (Part 3 of 6)

Categories: General Data Acquisition, General Data Logger, Thermocouple Data Acquisition Leave a comment

(Note: A consolidated version of this topic that contains all six part can be found here)

Effects of Common Mode Voltage Frequency

We’ve covered evaluating noise magnitude to see if noise elimination is worth the time and effort, and a procedure to test the instrument for proper operation.  In part 2, the concept of a common mode voltage (CMV) was introduced, and how a CMV can disrupt the operation of even isolated instruments if it’s too large.  Here I’ll continue the discussion of common mode voltages in isolated instruments, focusing not on its magnitude, but on its frequency.

Most of the CMV situations you’ll encounter revolve around 50/60 Hz power, or maybe 100/120 Hz in the presence of florescent lighting. These are low frequency CMVs that won’t interfere with the ability of an isolation barrier to perform. There are applications however that can wreak havoc on an isolation barrier; in the presence of high frequency noise generated by a welding machine or some other piece of industrial equipment, for example.

As the frequency of the CMV increases, the effectiveness of the isolation barrier decreases. This is due to the parasitic capacitance inherent to any non-optical isolation amplifier (parasitic capacitance). The formula used to calculate capacitive reactance (Xc) illustrates this relationship.

Think of capacitive reactance as the amplifier’s ability to resist noise. As the frequency (f) of the CMV rises, the isolation amplifier becomes less effective. At a certain threshold frequency the CMV is no longer blocked by the isolation barrier, but effectively coupled through it because of the barrier’s inherent parasitic capacitance.

 

How To Eliminate Noise in Data Logger and Data Acquisition Measurements (Part 2 of 6)

Categories: General Data Acquisition, General Data Logger, Thermocouple Data Acquisition Leave a comment

(Note: A consolidated version of this topic that contains all six part can be found here)

Common Mode Voltages and Isolated Instruments

In the premier edition of this series I talked about how you should initially approach noise elimination. The first step is to ensure that noise is a significant percentage of the full scale range of the analog channel, and then I explained a simple test to ensure that the instrument is not the noise source. If you’ve gotten this far I can assume that signal noise is significant, and the instrument is not defective.

Your next step is to look up the isolation specification of the instrument. Note that if you are under the impression that an instrument with differential analog inputs means that it is also isolated, you need to read this application note before proceeding. Differential and isolated analog inputs are not synonymous, and unless the specifications for the instrument state clearly that its inputs are isolated it most likely is not, regardless of its single-ended and/or differential configuration options. This edition of the Noise Elimination series assumes that you have an isolated instrument. I’ll deal with non-isolated instruments in later editions, since the subject becomes more complex in this situation.

How To Measure Common Mode Voltage Magnitude

So, you have read its specifications and determined that you have an isolated instrument. That’s great because noise-inducing  common mode voltages  (CMVs) can be largely, but not completely ignored. To absolutely verify that a common mode voltage is not a problem start by answering these questions:

  1. Is your field measurement one that virtually guarantees that a CMV is present? Examples are exposed thermocouple junctions that are attached directly to a conductor (powered or not), or the measurement of AC or DC current using a current shunt.
  2. If you answered YES to (1), then determine the magnitude of the CMV. With all field wiring disconnected except power and/or computer interfaces, use a battery-powered multimeter to measure the voltage from the field-side shunt or exposed conductive surface to the instrument’s chassis ground (not its low-side signal input.) Do this on the meter’s AC and DC ranges. Jot the value down and compare it with the isolation breakdown specification for the instrument.

If the measured value exceeds this spec then you have found the source of your noise, and you’re fortunate that the instrument still functions. Your two options are to lower the CMV if possible, or replace the instrument with one that provides a higher isolation spec.

In the next edition I’ll discuss CMV frequency and how this can destroy the best instrument isolation barriers.

 

How To Eliminate Noise in Data Logger and Data Acquisition Measurements (Part 1 of 6)

Categories: General Data Acquisition, General Data Logger, Thermocouple Data Acquisition Leave a comment

(Note: A consolidated version of this topic that contains all six part can be found here)

This happens almost daily: A customer contacts our support group to complain of noise on his measurement. Almost invariably a comparison is made between the output of his PC-based product and a connected DVM. The DVM shows a stable reading while the PC-connected measurement “is jumping all over the place.”

While there’s a slim chance that the PC-based instrument is to blame, that can only be true if the front end has been damaged in some manner. While not perfect, analog front ends on data acquisition and logger products are designed to inject only very small amounts of noise as a percentage of full scale measurement range.

What should you do if your setup exhibits a large amount of noise, so much so that it’s difficult to discern the true measured value? This series makes some suggestions that will help you identify and eliminate the problem. In this first part we’ll show you how to ensure that the instrument is functioning properly and is not the source of the noise. In subsequent posts we’ll dig progressively deeper into field wiring to expose some common problems that can manifest themselves.

So, is it the instrument, or not? Follow these steps to find out:

  1. Are you over reacting? Our software allows you to expand signal amplitude ranges to such a degree that a relatively small amount of noise can look overbearing. It’s a lot like cringing in horror as you look at a tick under a microscope. Read How To Put Signal Noise Into Perspective for guidance.
  2. Assuming that the noise you see is a significant percentage of the full scale range as determined by step (1), then your next move is to eliminate the instrument as the culprit. Disconnect all field wiring from it and simply short the analog inputs to common. I know. In the electronics world the word “short” implies something that’s very bad, but in this case it’s perfectly kosher. If your instrument has a single-ended configuration, short the (+) input to signal common. If it’s a differential configuration, connect the (+) input to the (-) and then connect those to signal common (if available, which it may not be for some differential configurations.) What you’re doing is establishing a firm zero volts on the front end of the channel so you can examine the instrument’s noise floor. In this configuration you should see no more than a few counts of noise around zero volts, dramatically and substantially less than you saw with a signal source connected, and roughly consistent with the manufacturer’s specification for the instrument’s noise figure. If you see anything else, then contact tech support for guidance because the chances are great that your instrument is defective.

Next time, we’ll delve into common field wiring issues that can and do produce more noise than a heavy metal concert.