Values in WinDaq are not correct when using DI-8B42-01 ModulesOn this PagePlace a 0.02µF capacitor across In+ and Ex+ on channels containing DI-8B42-01 modules SymptomsWinDaq displays values that are not consistent with the input on channels containing DI-8B42-01 modules. CauseA design issue involving the DI-8B42 module that can cause the amplifier to oscillate when used with certain types of transmitters. ResolutionTo resolve these issues follow the steps below. Place a 0.02µF capacitor across In+ and Ex+ on channels containing DI-8B42-01 modules.Place the capacitor as shown below:
Applies ToDI-718B(x), DI-788 and DI-78B input expanders containing DI-8B42-01 modules. |