DATAQ Logo site map | search

Low-Cost Data Logger Starter Kit

Click Here

DI-8B47 Linearized Thermocouple
Input Modules

Types J, K, and T

DI-8B Industrial Signal Conditioning Data Acquisition Module

 
  • Interfaces to Thermocouple Types J, K and T
  • Linearizes Thermocouple Signal
  • High Level Voltage Outputs
  • 1500Vrms Transformer Isolation
  • ANSI/IEEE C37.90.1
  • Input Protected to 240VAC Continuous
  • 120dB CMR
  • 70dB NMR at 60Hz
  • Low Drift with Ambient Temperature
  • Accurate CJC -40°C to +85°C
  • CSA, FM, and CE Certifications Pending
  • Mix and Match Modules
Part Number Type Input Range Accuracy* Price
(US)
DATAQ Store
DI-8B47J-01 J 0°C to +760°C
(+32°F to +1400°F)
±1.82°C $102
Quantity:
DI-8B47J-02 J -100°C to +300°C
(-148°F to +572°F)
±0.96°C $102
Quantity:
DI-8B47J-03 J 0°C to +500°C
(+32°F to +932°F)
±1.05°C $102
Quantity:
DI-8B47J-12 J -100°C to +760°C
(-148°F to +1400°F)
±2.10°C $102
Quantity:
DI-8B47K-04 K 0°C to +1000°C
(+32°F to +1832°F)
±2.40°C $102
Quantity:
DI-8B47K-05 K 0°C to +500°C
(+32°F to +932°F)
±1.05°C $102
Quantity:
DI-8B47K-13 K -100°C to +1350°C
(-148°F to +2462°F)
±3.60°C $102
Quantity:
DI-8B47K-14 K 0°C to +1200°C
(+32°F to +2192°F)
±2.88°C $102
Quantity:
DI-8B47T-06 T -100°C to +400°C
(-148°F to +752°F)
±2.40°C $102
Quantity:
DI-8B47T-07 T 0°C to +200°C
(+32°F to +392°F)
±0.75°C $102
Quantity:
*Includes conformity, hysteresis, and repeatability. Does not include CJC accuracy.
Note: Pricing outside the U.S. will be higher to account for shipping, taxes, duty, currency fluctuation, local warranty and technical support in your country.
DI-8B Signal Conditioners have typical lead times of 2 weeks ARO.

DI-8B47 Highlights

DI-8B modules are an optimal solution for monitoring real-world process signals and providing high level signals to a data acquisition system. Each DI-8B47 module isolates, filters, amplifies, and linearizes a single channel of temperature input from a thermocouple and provides an analog voltage output.

Linearization is accomplished using a four breakpoint piecewise linear approximation.

The DI-8B47 can interface to industry standard thermocouple type J, K, and T and has an output signal of 0V to +5V. Each module is cold-junction compensated to correct for parasitic thermocouples formed by the thermocouple wire and screw terminals on the mounting backpanel. Upscale open thermocouple detect is provided by an internal pull-up resistor.

Signal filtering is accomplished with a three-pole filter optimized for time and frequency response which provides 70dB of normal-mode-rejection at 60Hz. One pole of this filter is on the field side of the isolation barrier for anti-aliasing, and the other two are on the system side.

A special input circuit on the DI-8B47 module provides protection against accidental connection of power-line voltages up to 240VAC.

The modules are designed for installation in Class I, Division 2 hazardous locations and have a high level of immunity to environmental noise.


Want to port your data into a spreadsheet or a database?

Check out our add-on software:

Port Data Acquired with your DI-710 Data Logger to Excel

WinDaqXL This WinDaq Software data acquisition add-on allows you to port acquired data directly into a Microsoft Excel Spreadsheet in real time. 60-day Trial Version available.

Also check out the DATAQ Developer Network to learn how to use our ActiveX controls with Microsoft Excel and Access.

Port Data Acquired with your DI-710 Data Logger to MySQL

WinDaq/MySQL This free WinDaq Software data acquisition add-on allows you to port acquired data directly into a MySQL database in real time.

         
 

Data Acquisition Product Links:

(click on text to jump to page)

mail Webmaster