Cannot change direction of D1 bit to Output (stays as an input)On This PageSymptomsD1 (Remote Events) is always set to Input and cannot be changed to an Output for PC-tethered USB DI-710s and DI-715Bs. CauseBy default, the remote event flag is enabled, thus D1 is set to input. The direction of the digital bit is overruled by the remote event flag. Since the PC-thethered instruments have no access to the stand-alone configuration panel (where it can be changed in stand-alone units), we can’t change D1 to an output. ResolutionResolution requires you to run a separate utility available from DATAQ Instruments Technical Support (or via hyperlink in this article). Before continuing, close the DATAQ Instruments Hardware Manager program.
You can now change the second bit (D1) from an Input to an Output using the Digital I/O Setup command in the Hardware Manager drop-down command menu. Applies ToThe information in this article applies to: DATAQ Instruments Hardware Manager. DI-710 and DI-715B PC-tethered USB Products.
|
||||||||||||
|
||||||||||||